Choosing the right sensor
Which sensor suits my measuring site?
What matters is the measured variable, installation location, maximum water column and medium. For the water level there is the Aquatos Radar from above, the piezo pressure sensor for wells and deep levels and the ceramic pressure sensor for basins and wastewater plants. Describe your site to us and we will get back to you with a recommendation.
Radar or pressure probe?
The radar sensor measures from above and needs a clear view of the water surface. A pressure probe measures directly in the water, for example in wells and gauge tubes. What matters is the installation location, maximum water column and medium.
Piezo or ceramic pressure sensor?
The piezo pressure sensor Type 0312 covers 5 to 100 m H₂O and reaches ±0.15 % of full scale, intended for groundwater and deep levels. The ceramic pressure sensor Type 0420 measures 0 to 1 bar, about 10 m H₂O, and is intended for basins and wastewater plants. We agree the media compatibility with you in advance.
Which variables do the sensors cover?
Water level and fill level, conductivity, temperature, oxygen, redox and nitrate, with multiparameter probes also pH and turbidity, plus precipitation, air temperature and humidity. We put together further single sensors such as ammonium, phosphate, free chlorine or salinity to suit the measuring task.
Interfaces and setup
What is SDI-12?
A manufacturer-independent digital bus for environmental sensors. Several sensors share one cable and the logger queries them one after the other. This saves power and allows cable lengths over 500 m.
SDI-12 or Modbus?
SDI-12 is made for battery-powered environmental sensors. If a sensor is available in both variants, choose SDI-12. Modbus is right when an existing industrial sensor needs to be integrated. The converter Type 0210 takes care of that.
Can sensors from other manufacturers be connected?
Yes. SDI-12 compliant sensors run directly on the bus, Modbus sensors via Type 0210, pulse outputs from water meters and tipping buckets via Type 0330.
How are the sensors set up?
On site via Bluetooth directly in the browser, without installing an app. For the radar, the raw scan and live plot help with alignment; on the piezo probe, the tare function adjusts the zero point.
How are the sensors powered?
The Aquatos Radar runs on 3.6 to 16 V DC, the pressure probes on 2.8 to 14 V and the adapters Type 0330 and Type 0210 on 2.8 to 16 V. Quiescent current is below 10 µA for the pressure probes and adapters and below 30 µA at 4 V for the radar.
Which data loggers fit?
SDI-12 compliant loggers, for example the Aquatos loggers from TerraTransfer GmbH: Aquatos Web LTX with cellular connection and up to 24 channels, Aquatos mini with Bluetooth read-out and Aquatos nano as a 1-inch submersible with logger.
Pressure probes
Gauge or absolute pressure?
A pressure probe in the water measures the water column plus air pressure. The gauge pressure probe brings the air pressure to the diaphragm through a capillary in the cable, so the reading is the water level directly. The absolute pressure probe needs no capillary, but the air pressure must be recorded separately and subtracted. Otherwise, weather-related fluctuations quickly amount to 10 to 30 cm of water column.
What should I watch out for during installation?
Tube diameter at least 1 inch, diaphragm facing down. Lay the cable without kinks, with a bending radius of at least 15 times the cable diameter. From 50 m cable length, add a steel rope for strain relief. Keep the capillary end at the logger dry and open.
How often must calibration be done?
The probes leave the factory calibrated. For critical measurements we recommend a field calibration of the piezo probe against a known water level every 2 to 3 years.
Is there a solution for salt water?
Yes. For saline waters as well as coasts and tidal zones, the piezo pressure sensor is available with a titanium diaphragm.
Radar
How is the Aquatos Radar mounted?
Vertically above the water surface on a bridge, wall or outrigger, with a clear view of the water surface. The distance to the medium should be at least 0.25 m. The sensor is not intended for permanent underwater installation.
What about interfering echoes, for example in shafts?
The radar lens focuses the beam to about 10°. Up to three echoes are recorded with their signal strength, and interfering reflections can be identified and masked out using measuring range, sensitivity and sorting.
Water quality, temperature and precipitation
Single sensor or multiparameter probe?
Single sensors target one variable. Multiparameter probes record several together at one measuring point, for example pH, oxygen, conductivity and turbidity. The choice depends on the water body, measuring task and installation.
Does nitrate measurement need reagents?
No. The nitrate probe measures with UV light directly in the water, without reagents or sampling.
How many measuring points does a thermistor chain have?
Up to 30 nodes on one chain, with cable lengths up to 200 m. You decide on the positions.
Tipping bucket, weighing gauge or combination sensor?
The heated tipping bucket measures all year round using a funnel and tipping mechanism. The weighing gauge records to WMO-No.8, including snow and hail, and we recommend a wind shield and antifreeze solution in winter. The WS100 detects intensity and type of precipitation, but is no replacement where exact amounts are required.
Ordering and service
Where can I order?
You will find sensors and accessories in the TerraTransfer shop. For special versions or if you would like advice, just send us an enquiry.
Which details help with an enquiry?
Measured variable, measuring range and installation, and for aggressive media also the medium.
Where can I find datasheets and manuals?
On the respective sensor page, where documents are available. We are happy to send you anything else.
Who do I contact if there is a defect?
Please get in touch using our contact details. Repairs are carried out at the manufacturer.
Can I also get a complete monitoring system?
Yes, TerraTransfer GmbH takes care of the complete monitoring system with logger, installation and operation. TerraTransfer Scientific focuses on the sensors.
Anything else?
Describe the measured variable, measuring range and installation. We will get back to you with a recommendation.